A quantitative method of capillary electrophoresis with sample stacking induced by moving reaction boundary (MRB) was developed for
sensitive determination of oxymatrine (OMT) and matrine (MT) in rat plasma. The experimental conditions were optimized firstly. Below are
the optimized experimental conditions: 20mM sodium formate solution (HCOONa, adjusted to pH 10.70 by ammonia) as sample solution, 3 min
14 mbar sample injection, 40mM formic buffer (HCOOHHCOONa, pH 2.60) as stacking buffer, 7 min 14 mbar injection of stacking buffer,
100mM HCOOHHCOONa (pH 4.80) as separation buffer, 73 cm capillary (effective length 64 cm), 21 kV voltage, 210 nm wavelength. Under
the optimized conditions, higher than 60-fold sensitivity improvement of the stacking was simply achieved as compared with capillary zone
electrophoresis, and the detectable limits obtained for OMT and MT were 0.26 and 0.19 gmL−1, respectively. Then, numerous demonstrations were carefully performed for the methodological validations of OMT and MT in rate plasma, including high specificity of method, good linearity (r = 0.9993 for OMT, r = 0.9991 for MT), fair wide linear concentration range (1.3065.00 gmL−1 for OMT, 0.8442.00 gmL−1 for MT), low limit of detection (1.03 gmL−1 for OMT, 0.38 gmL−1 for MT), less than 5% intra- and inter-day variance value, and higher than 96% recovery of OMT and MT in plasma. The developed method could be used for the trace analyses of OMT and MT in plasma and was finally used for the investigation on pharmacokinetic study of OMT in rat plasma. GX毛细管电泳仪(紫外检测)CL1020
相关产品