仪器网(yiqi.com)欢迎您!

| 注册2 登录
网站首页-资讯-专题- 微头条-话题-产品- 品牌库-搜索-供应商- 展会-招标-采购- 社区-知识-技术-资料库-方案-产品库- 视频

应用方案

仪器网/ 应用方案/ 当温湿度控制器显示湿度异常的状况,怎么办呢?

立即扫码咨询

联系方式:13790142908

联系我们时请说明在仪器网(www.yiqi.com)上看到的!

扫    码    分   享

【概述Summarize】当温湿度控制器显示湿度异常的状况,该如何辨别以及解决?

When the temperature humidity controller shows the condition of abnormal humidity, how ?


恒温箱主要用于测试和确定电工、电子及其他产品及材料进行高温、低温、交变湿热度或恒定试验的温度环境变化后的参数及性能。Thermostat is main used for testing and determining the parameters and performance of electrical and electronic and other products and materials after the temperature environment changes in high and low temperature alternating humidity and heat degree or constant test.


【检查步骤 Check Steps

  1. 查看控制机房的欠水超温保护器,是否设定在150℃的位置。

    Check whether the under-water and over-temperature protector in the control unit is set at the position of 150 Celsius.

  2. 查看机房内,管路是否过脏,造成水流不顺。

    Check the unit, whether the pipeline is too dirty, causing water flow.

  3. 查看上水箱是否有水,如果没有水,再查看下水箱是否有水。

    Check whether has water in top water tank, if not, same to check the bottom water tank.

  4. 查看是否湿度部份的固态继电器烧毁短路: 如果加热器未烧毁,使用三用电表交流电压档,电压档位开到600伏特的位置,将红黑棒分别放在线号标注为H的那一颗固态继电器交流两侧,将湿度部份的湿度设定值设置0%,此时湿度部份的固态继电器的指示灯不会亮起,如果量测的电压值没有变化,维持在10V以下代表固态继电器烧毁呈现短路状态。或者首先将湿度部份设定值设为0%,再看机房内加湿桶是否水在煮沸状态。

    Check if the humidity of solid state relay burned short-circuit: if heater is not burning, used three meters ac voltage, the position of the gear drive voltage to 600 volts, will be red and black on the annotation of line number for H that a solid state relay communication on both sides, set the humidity of humidity value 0%, humidity at this time of the light on the solid state relay can't light up, if the measured voltage value did not change, keep under 10 v represents the solid state relay burned present state of short circuit Or the first set of 0% humidity part of the set value, then see whether humidifying barrels of water inside the room in the boiling state.


【解决办法 Solutions

  1. 1、将欠水超温保护器转至150℃的位置。

    Turn under-water over-temperature protector to position 150 Celsius.

  2. 2、首先将机台后方总电源关上,然后将加湿桶下方的排水阀,打开到与管路平行的位置,带上水箱的水都排光的时候,利用板手将加湿桶打开,将内部因长期做湿度的水垢清除,然后将硅胶管路拔除,在水中搓洗。保养完毕后按照拆除的方式,反方向装回去。装置完毕后,需将加湿桶下方的排水阀,关闭到与管路垂直的位置,然后再将机台后方总电源打开。

    Firstly to turn off the total power supply behind the machine, then open the drainage valve under the humidifying barrel, till to a position parallel to the pipe line. Water shortage from top water tank, open the humidifying barrel by the spanner, scale removal and silica gel tube was removed, clean in water. After the maintenance in accordance with the demolition mode, put it back in the opposite direction, after installation, turn on the drainage valve with the pipeline on the vertical position, then turn on the total power supply.

  3. 3、下水箱都无水,将下水箱加水,即可解决。上水箱如果没有水,下水箱有水,则为下水箱打水马达坏掉,通知皓天黄工维修客服部门上门进行维修; 

    There is no water in the bottom tank, will add water in that tank, solve it ! If there is no water in the top tank, there is water in bottom tank, the bottom tank to drain water, means the motor broken. So inform the maintenance customer service department of Haotian company to conduct on-site.

  4. 4、更换湿度部份的固态继电器(线号标注为H的那一颗固态继电器),或者通知黄工售后上门进行维修。

    Replace the solid state relay of the humidity part (the one marked with line No. H), or notify the after sales staff to conduct on-site maintenance.


相关产品

参与评论

全部评论(0条)

推荐方案

当温湿度控制器显示湿度异常的状况,怎么办呢?
当温湿度控制器显示温度异常的状况,皓天来帮您
高低温交变湿热试验箱温湿度控制器显示湿度异常的故障及解决办法
温湿度控制器显示异常,按钮点击无动作
恒温恒湿试验箱温湿度控制器显示异常的解决方案
温湿度控制器显示温度异常(超温报警)的故障及解决方法
高低温交变湿热试验箱温湿度控制器显示压力异常的故障及解决办法
控制器显示缺水报警(水位异常)的故障怎么解决?
当温度采集仪上测得的温度与控制器上显示的差距大
应对恒温恒湿试验箱温湿度控制器显示故障的全面指南
可程式恒温恒湿箱湿度突然显示100%
恒温恒湿试验箱湿度降不下去怎么办?
恒温恒湿试验箱湿度降不下去可能是由多种原因引起的,需要逐步排查和解决。以下是常见原因及解决方法:
恒温恒湿试验箱控制器出现故障了怎么办
当恒温恒湿的湿度上不去的时候怎么解决
当恒温恒湿设备的湿度无法达到设定值时,可以采取以下措施
恒温恒湿试验箱无法操作控制该怎么办呢?
小型模拟环境试验设备温湿度显示的破解之道
基于DSP的数字化温湿度智能控制器设计.pdf
恒温恒湿试验箱的湿度无法达到设定值该怎么办?
恒温恒湿试验箱湿度无法达到设定值,可从以下几个方面进行排查和解决:
恒温恒湿试验箱温湿度控制器出现异常解决方法
高低温试验箱在使用中冰堵了该怎么办呢?

在线留言

上传文档或图片,大小不超过10M
换一张?
取消