The separation of acyclovir (ACV) by high performance capillary electrophoresis (HPCE) with on-column amperometric detection using -amino-5-mercapto-3,4-dithiazole (AMD) as internal standard is described. The calibration line was linear in the range of 0.5 - 20 mg/L of ACV. The detection limit was 0.15 mg/L of ACV. Its recovery ranged from 98 to 101% with relative standard deviations (RSDs) from 1.9 to 3.2% (n = 5). This method was successfully used for determining ACV in some pharmaceuticals and human urine. Comparable results with HPCE with ultraviolet (UV) detection and amperometric detection were obtained. GX毛细管电泳仪(紫外检测)CL1030