ABSTRACT
摘 要
01
建立环孢素眼用乳剂体外释放方法。
To develop a method for characterizing in vitro release of cyclosporine ophthalmic emulsion.
02
采用Franz扩散池,聚偏氟乙烯膜滤膜,以缓冲盐-无水乙醇(40:60)为接收液,时间点为60、125、190、255、320、385min。
The Franz diffusion cell and the polyvinylidene fluoride membrane were adopted with buffer-ethanol (40:60) as receiving media. The sampling time was set at 60,125,190,255,320,385min, respectively.
03
方法学验证表明,体外释放方法的滤膜惰性、专属性、灵敏度、选择性均符合规定。测定方法的定量限为0.07μg·mL-1,在0.07~44.62μg·mL-1范围呈现良好的线性关系,回收率为98.9%。依据FDA的判定原则,自研制剂与参比制剂体外释放行为一致。
The in vitro release method showed that the inertia of membrane, specificity, sensitivity and selectivity met the requirements. The validation of HPLC showed that the quantitative limit of the method was 0.07μg·mL-1, and a good linear relationship between the concentration range of 0.07-44.62μg·mL-1. The average recovery was 98.9%. Compared with the original preparation by FDA guideline, the in vitro release of the self-developed preparation was the same as the reference preparation.
04
本法适合环孢素眼用乳剂的体外释放度评价。
This method is suitable for the invitro release evaluation of cyclosporine eye drops.
Key words
关 键 词
环孢素眼用乳剂;体外释放(IVRT);参比制剂;自研制剂;方法学验证
cyclosporine ophthalmic emulsion; in vitro release (IVRT); reference preparation; self-developed preparation; method validation
CONTENT
内 容
点击下方链接查看全文
LOGAN SINCE 1990
客服热线
Tel +400-1003-718全部评论(0条)
LOGAN 半自动水平式吸入制剂多级撞击器系统NGI System
报价:面议 已咨询 1485次
LOGAN 12位溶出自动取样系统 SYSTEM 860DL
报价:面议 已咨询 3382次
LOGAN UDT-818A-8 8位溶出度仪
报价:面议 已咨询 5633次
LOGAN 硬度仪 HDT-500L
报价:面议 已咨询 1430次
LOGAN 8位溶出渗透高效液相系统 PERMETRO 800 HPLC
报价:面议 已咨询 1413次
LOGAN 溶出杯自动清洗系统 AVC-100HTP
报价:面议 已咨询 1435次
LOGAN 8位溶出自动取样系统 SYSTEM 850DLA
报价:面议 已咨询 1445次
CULTURED SKIN人工培养皮肤
报价:面议 已咨询 1414次
①本文由仪器网入驻的作者或注册的会员撰写并发布,观点仅代表作者本人,不代表仪器网立场。若内容侵犯到您的合法权益,请及时告诉,我们立即通知作者,并马上删除。
②凡本网注明"来源:仪器网"的所有作品,版权均属于仪器网,转载时须经本网同意,并请注明仪器网(www.yiqi.com)。
③本网转载并注明来源的作品,目的在于传递更多信息,并不代表本网赞同其观点或证实其内容的真实性,不承担此类作品侵权行为的直接责任及连带责任。其他媒体、网站或个人从本网转载时,必须保留本网注明的作品来源,并自负版权等法律责任。
④若本站内容侵犯到您的合法权益,请及时告诉,我们马上修改或删除。邮箱:hezou_yiqi
参与评论
登录后参与评论